제품소개

RTD (Temperature sensor / DC)

· Function(기능)
RTD Temperature Sensor Converter (isolated)
· Input(입력)
RTD temperature sensor
· Output(출력)
mA,V / dc (isolated)
· Other Spec.(기타)
제품 상세정보

The Model RTD accepts 2-or-3-wire RTD inputs and provides isolated and linearized DC outputs.
Linearization and upscale burnout protection standard; downscale or no burnout optional.

ORDERING INFORMATION
Specify code number and variables. Burnout is set to up-scale.
Specify when downscale or no burnout protection is required
-Code no.: RTD-Model/Input/Output1/Ouput2-Power
(Ex : RTD-4C8N-X)
-Temperature Range (e.g.0~100 C)
-Special Output Range (Ex.0~20mA)

GENERAL SPECIFICATIONS
-Housing Material : Plastic ABS(Polycarbonate) Black
-Isolation : Input to Output to Power
-Adjustment : Zero & Span(Separation) ±10%of F.S
-Burnout Protection : Upscale standard (Downdscale option)
-Display : 4 digits

INSTALLATION
-Power Supply : AC 100~120V or 200~230V 50/60Hz (about 3VA),DC 18~30V (about 3Watts)
-Operating Temperature : -5 ~ +55 ℃
-Operating Humidity : 90% RH max.(Non-condensing)
-Mounting : Wall or Rail mounting (Plug-in Type)
-Dimension : 52(W)×77(H)×112(D) (Unit:mm)
-Weight : about 400g

PERFORMANCE
-Accuracy : ±0.2%of Full Scale
-Temperature Coefficient : ±0.015%/℃
-Response Time : 0.5 Sec or less(0~90%),Normal
-Input Range : 100Ω~ 100KΩ (Fee : 3-Wire only)
-Insulation Resistance : 100MΩ or more with 500V DC between Input/Output/Power
-Dielectric Strength : 1500V AC 0.5mA/Min (Input to Output to Power)
-Display Range : -1999 to 9999 (Available Decimal Point,Unit: C)
-Input Sensing : 2mA DC (Pt 100Ω RTD Sensor)

MODEL & SUFFIX CODE
Model Type Selection
2 : 1 Output with Display
4 : 1 Output without Display
7 : 2 Outputs with Display
9 : 2 Outputs without Display

Input Type & Range Selection
A : -50 ~ +50 ℃ for JIS spec. B : 0 ~ 50 ℃ for JIS spec.
C : 0 ~ 100 ℃ for JIS spec. D : 0 ~ 200 ℃ for JIS spec.
E : 0 ~ 300 ℃ for JIS spec. F : 0 ~ 400 ℃ for JIS spec.
G : -50 ~ +50 ℃ for DIN spec. H : 0 ~ 50 ℃ for Din spec.
I : 0 ~ 100 ℃ for DIN spec. J : 0 ~ 200 ℃ for DIN spec.
K : 0 ~ 300 ℃ for DIN spec. L : 0 ~ 400 ℃ for DIN spec.
R : Other Special Spec.

Output Type & Range Selection (Available for Output 1 & Output 2)
1 : DC 0~10mV (Load Resistance : 10 KΩ or more)
2 : DC 0~100mV (Load Resistance : 100 KΩ or more)
3 : DC 0~5V (Load Resistance : 5 KΩ or more)
4 : DC 0~10V (Load Resistance : 10 KΩ or more)
5 : DC 1~5V (Load Resistance : 5 KΩ or more)
6 : DC 0~1mA (Load Resistance : 0~15 KΩ)
7 : DC 0~20mA (Load Resistance : 0~750 Ω)
8 : DC 4~20mA (Load Resistance : 0~750 Ω)
R : Other Special Spec.
N : Not Used

Power Supply
W : AC 200~230V for 50/60 Hz
X : AC 100~120V for 50/60 Hz
Y : DC 18~30V
Z : Other Special Spec.

tel: 01087315801